Comparative analysis of two retinal fractures with ultrabroad‑angle fundus photography systems

  • Authors:
    • Yong Guo
    • Hong Yan
    • Chenjun Guo
    • Jue Wang
  • View Affiliations

  • Published online on: January 4, 2024     https://doi.org/10.3892/br.2024.1722
  • Article Number: 34
  • Copyright: © Guo et al. This is an open access article distributed under the terms of Creative Commons Attribution License.

Metrics: Total Views: 0 (Spandidos Publications: | PMC Statistics: )
Total PDF Downloads: 0 (Spandidos Publications: | PMC Statistics: )


Abstract

The aim of the present study was to compare the performance of the Opel Panorama 200 and Zeiss Clarus 500 (Carl Zeiss AG) systems in diagnosing retinal fractures. Human subjects were selected from 298 fundus examinations (531 eyes) in ophthalmology from February 2021 to June 2021, including 68 patients with retinal fissures (95 eyes). All fundus tests were performed with Opel Panoramic 200. Zeiss Clarus 500 (Carl Zeiss AG) fundus photography, slit‑lamp full retinal lens (Ocular Mainster Wide Field; Ocular Instruments), and retinal laser photocoagulation was performed for all affected eyes. The diagnostic sensitivity of the two examination methods was compared, and their sensitivities for posterior retina, peripheral nose, crystal eye, cataract, positive experiment, and myopia testing were compared. In all, 68 patients (95 eyes) were clinically examined and treated 112 laser times. For retinal fractures, the Opel Panorama 200 used a check sensitivity of 89.5%, and the Clarus 500 check had a sensitivity of 94.7%, with the difference being non‑significant (P=0.358). Moreover, Clarus 500 diagnosed the sensitivity of the temporal periphery significantly higher than that of Opel Panorama 200 (P=0.048). Opel Panorama 200 displayed statistically significant sensitivity compared with Clarus 500 diagnosis with crystalline and crystal fewer eyes (P>0.05); Clarus 500 sensitivity for cataract diagnosis (crystal turbidity level 3 and above) was significantly higher than that of Opel Panorama 200 (P=0.033). Opel Panoramic 200 displayed significant sensitivity to ocular myopia and medium to moderate myopia (P>0.05). Clarus 500 diagnosed high myopia with a significantly higher sensitivity than Opel Panorama 200 (P=0.045). Opel Panorama 200 and Zeiss Clarus 500 displayed the same level of sensitivity to retinal fissures, with improved sensitivity in refractive turbidity and for retinal fissures located in the far periphery of the temporal side.

Introduction

Comprehensive and objective information acquisition and analysis of the fundus and anterior segment of the eye are particularly important for the early diagnosis of pathological changes, such as retinal tears, degeneration, hemorrhage and detachment (1). The detection of anterior segment-related diseases, such as cataracts and myopia, provides an important basis for clinical treatment decisions (2). In the past decade, retinal imaging technology has developed rapidly, significantly improving the probability of early diagnosis and treatment of fundus diseases and improving the prognosis and visual function of patients. In particular, the emergence of ultrawide-angle camera technology enables simultaneous imaging of the posterior pole and peripheral regions of the retina in one image, which simplifies the diagnosis process of fundus diseases (3). The first ultrawide-angle camera system used in ophthalmology clinics is the Optos Panorama 200 image system (Optos), which has a sensitivity rate of 70-80% for peripheral retinal microscopic holes (4). The just-launched Zeiss Clarus 500 (Carl Zeiss AG) imaging system (Carl Zeiss Meditech AG) ultra-wide-angle camera system has superior advantages; however, there remains no comparative study of these two ultrawide-angle imaging systems (5). The present study compared the sensitivity of the Auberg Panorama 200 and the Zeiss Clarus 500 for microscopic tears in the peripheral retina, as well as the sensitivity of retinal tears under aphakia, lens opacity, or intraocular lens to explore the clinical role and significance of the ultrawide-angle camera system and to provide clinical evidence and new diagnostic ideas for the diagnosis of peripheral retinal diseases.

Materials and methods

Research objects

The present study was a prospective observational study of 298 patients, including 174 men and 124 women aged between 8 and 90 years old. A total of 531 eyes were examined at Xi'an Purui Eye Hospital (Xi'an, China) from February 2021 to June 2021. Of these, 68 patients (95 eyes) were diagnosed with retinal tears. Slit-lamp examinations were performed in all eyes, and patients with keratopathy or previous ocular trauma or ocular surgery, and intraoperative surgical complications were excluded. All patients underwent optometry, intraocular pressure, and visual acuity examinations, and they all voluntarily participated in this study and signed informed consent forms. The present study (approval no. XPR-2021-0007) was approved by the Xi'an Bright Eye Hospital Ethics Committee (Xi'an, China).

Inspection method

The conjunctiva, cornea, anterior chamber, pupil, and lens of all eyes were examined by the same deputy chief physician of the ophthalmology clinic using a Zeiss SL30 slit microscope (Carl Zeiss AG). Midorie eye liquid was used to dilate pupils one time for 5 min, three times in total. The pupils were dilated to 8 mm in diameter, and images of the affected eyes were captured with Opel Panorama 200 and Zeiss Clarus 500 (Carl Zeiss AG). The examination results were analyzed by two senior attending physicians with fundus specialty. All eyes were examined with slit-lamp pan retinoscopy (Ocular Mainster Wide Field; Ocular Instruments). All eyes with retinal tears were treated with retinal laser photocoagulation.

Observation indicators

The aim of the present study was to compare the sensitivity and specificity of two ultrawide-angle fundus photography systems for the diagnosis of retinal tears and to compare the sensitivity of the two ultrawide-angle fundus photography systems in the diagnosis of the retinal posterior pole, temporal periphery, nasal periphery, phakic eyes, cataract, emmetropia, moderate to low myopia, and severe myopia. The following has to be noted: Sensitivity=number of true positive cases/(number of true positive cases + number of false negative cases) x100.0%. Specificity=number of true negative cases/(number of true negative cases + number of false positive cases) x100.0%.

Statistical analysis

Statistical analysis was performed using SPSS 13.0 software (SPSS, Inc.; IBM Corp). First, a normality test of the data was conducted, and the paired t-test was used for the data that displayed normal distribution. Non-parametric tests were used for data that did not meet the conditions, while all enumeration data were expressed as cases (percentages), and the McNemar's chi-square test was used for comparison between groups. P<0.05 was considered to indicate a statistically significant difference.

Results

Comparison of the value of the two systems in the diagnosis of retinal tears

The Opel Panorama 200 has a sensitivity of 89.50% and a specificity of 95.40% for the diagnosis of retinal tears; the Zeiss Clarus 500 (Carl Zeiss AG) system had a sensitivity of 94.70% and a specificity of 98.90% for the diagnosis of retinal tears. There was no statistically significant difference in the diagnostic sensitivity and specificity between the two systems (P>0.05). The details are shown in Table I.

Table I

Comparison of the diagnostic value of two systems for retinal tears.

Table I

Comparison of the diagnostic value of two systems for retinal tears.

 OpelZeiss
 PositiveNegativePositiveNegative
Pathological result    
     Positive8510905
     Negative174195431
Total10242995436
  Sensitivity Specificity
χ2 2.205 1.056
P-value 0.148 0.232
Comparison of the sensitivity of the two systems in diagnosing the temporal, nasal and posterior poles

The sensitivity of the Opel Panorama 200 in diagnosing the temporal side was 86.00%, and that of the Zeiss Clarus 500 system in diagnosing the temporal side was 93.00%; the sensitivity of the Opel Panorama 200 in diagnosing the nasal side was 93.10%, while that of the Zeiss Clarus 500 system in diagnosing the nasal side was 96.60%; and the sensitivity of the Opel Panorama 200 in diagnosing the posterior pole was 100.00%, and the sensitivity of the Zeiss Clarus 500 system in diagnosing the posterior pole was 100.00%. There was no significant difference between the two systems in diagnosing the nasal and posterior poles (P>0.05). The sensitivity of the Zeiss Clarus 500 system in diagnosing the temporal side was significantly higher than that of Opel Panorama 200 (P<0.05) (Table II).

Table II

Comparison of the sensitivity of the two systems for diagnosing the temporal, nasal, and posterior poles.

Table II

Comparison of the sensitivity of the two systems for diagnosing the temporal, nasal, and posterior poles.

 TemporalNasalPosterior poles
 OpelZeissOpelZeissOpelZeiss
True positive4953272899
False negative842100
Total5757292999
χ2 3.941 0.003 0.000
P-value 0.042 0.925 1.000
Comparison of the sensitivity of the two systems to retinal holes in aphakia and cataracts

The sensitivity of Opel Panorama 200 for diagnosing phakic eyes was 89.50%, and that of the Zeiss Clarus 500 system for diagnosing phakic eyes was 94.20%; the sensitivity of Opel Panorama 200 for diagnosing aphakia was 88.90%, and that of the Zeiss Clarus 500 system for diagnosing aphakia was 88.90%. The sensitivity of the Opel Panorama 200 in diagnosing cataract eyes was 80.60%, and that of the Zeiss Clarus 500 system in diagnosing cataract eyes was 91.70%. There was no significant difference between the two systems in diagnosing aphakia (P>0.05). The sensitivity of the Zeiss Clarus 500 system in diagnosing cataract eyes was significantly higher than that of Opel Panorama 200 (P<0.05) (Table III).

Table III

Comparison of the diagnostic sensitivity of the two systems in aphakia and cataracts.

Table III

Comparison of the diagnostic sensitivity of the two systems in aphakia and cataracts.

 Phakic eyesAphakiaCataract eye
 OpelZeissOpelZeissOpelZeiss
True positive7781892933
False negative951073
Total8686993636
χ2 0.869 0.565 4.205
P-value 0.554 0.651 0.044
Comparison of the diagnostic sensitivity of the two systems in refractive error

The sensitivity of the Opel Panorama 200 in diagnosing emmetropia was 92.90%, and that of the Zeiss Clarus 500 system in diagnosing emmetropia was 100.00%; the sensitivity of the Opel Panorama 200 in diagnosing moderate and low myopia was 91.60%, and the sensitivity of the Zeiss Clarus 500 system in diagnosing moderate and low myopia was 96.70%. The sensitivity of Opel Panorama 200 in diagnosing severe myopia was 80.10%, and the sensitivity of the Zeiss Clarus 500 system in diagnosing severe myopia was 85.70%. There was no significant difference in the sensitivity of the two systems in diagnosing emmetropia and moderate to low myopia (P>0.05). The sensitivity of the Zeiss Clarus 500 system in diagnosing severe myopia was significantly higher than that of the Opel Panorama 200 (P<0.05) (Table IV).

Table IV

Comparison of the sensitivity of the two systems in the diagnosis of refractive errors.

Table IV

Comparison of the sensitivity of the two systems in the diagnosis of refractive errors.

 EmmetropeLow to moderate myopiaSevere myopia
 OpelZeissOpelZeissOpelZeiss
True positive131455581718
False negative105243
Total141460602121
χ2 0.776 0.447 3.805
P-value 0.587 0.875 0.050

Discussion

Retinal tears in individuals with moderate to high refractive errors are still the main risk factors for rhegmatogenous retinal detachment. Because patients have only a flash of light or have no symptoms, it is difficult to diagnose, and the opportunity for treatment is missed. Retinal detachment in the later stage will endanger visual function, and the prognosis is not favorable. Therefore, early and accurate diagnosis and timely laser treatment are important means of reducing the harm caused by retinal tears (6,7). Handheld fundus cameras remian the main diagnostic method for fundus diseases worldwide. These cameras are widely used in the initial screening of fundus diseases because of their easy portability and simple operation. However, handheld fundus cameras have limitations like a narrow field of view, low resolution and potential for misdiagnosis (8,9). This is why pan-retinoscopy remains the primary diagnostic method for retinal tears. It has the advantages of a wide field of view, strong three-dimensional sense, deep depth and clear vision. This technology requires a long learning curve and is difficult to master and promote in a short time. Currently, it is only used as a means for surgeons to determine the diagnosis, select the surgical method, and estimate the prognosis of the surgery. Before the operation, the patient needs to be fully dilated and under topical anesthesia, and the operation requires favorable cooperation from the patient. The rough operation has discomfort and risks, all of which have prevented the widespread use of pan-retinoscopy (10). The ultrawide-angle fundus imaging system is a new fundus examination method emerging with the development of imaging technology. Its advantage is that it can obtain fundus image information up to more than 200˚ in the natural state with small pupils (11). It can display important structures, such as the retina and choroid, and can integrate them to form a comprehensive fundus color image, which can extract 80.0% of the retinal area image information. The ultrawide-angle fundus imaging system has been used in the diagnosis and follow-up of fundus diseases, such as diabetic retinopathy, due to its advantages of non-invasiveness, wide angles and non-mydriasis (12).

The Opel Panorama 200 is currently the main clinical ultrawide-angle fundus imaging system. The Virtual Point laser scanning technology used (13) covers >80.0% of the retina (14), and it distinctly shows the retina, choroid and other important structures of the fundus (12). The Zeiss Clarus 500 system is the world's first ultrawide-angle fundus imaging system that can simultaneously provide true fundus color and high-resolution images, with a single shooting range of 133˚ (15) and a 200˚ shooting range in ultrawide-angle mode, which is not only restoring the comprehensive and real information of the lesion to the greatest extent but also reduces the contact between doctors and patients through the non-contact mode and improves the efficiency of diagnosis and treatment (13). At present, two ultrawide-angle fundus imaging systems, the Opel Panorama 200 and Zeiss Clarus 500 systems, have certain applications in fundus diseases, especially diabetic retinal diseases (16); however there is a lack of systematic reports on their diagnostic effects in retinal tears. Falavarjani et al (17) found that in the examination of the sugar net, the retinal imaging under the Opel Panorama 200 imaging system was distorted, and the displayed hemorrhage range was larger than the actual hemorrhage range. Similar results were obtained in the present study. Both the Opel Panorama 200 and Zeiss Clarus 500 systems can obtain a 200˚ fundus range; however, the Opel Panorama 200 has a wide range of artifacts and distortions in peripheral retinal imaging. The Opel Panorama 200 misdiagnosed the condition as a small hemorrhage, while the Zeiss Clarus 500 system obtained clear peripheral retinal images. Price et al (18) considered that the Zeiss Clarus 500 system was superior to the Opel Panorama 200 system in the display of small diseases. However, Kernt et al (19) and Neubauer et al (20) both suggested that the color separation technology of Opel Panorama 200 could make up for these shortcomings. For the observation of rare diseases, such as children's fundus disease (21), sickle retinopathy (22) and familial amyloidosis (23), the two ultrawide-angle fundus imaging systems did not show significant differences. This might be caused by too few observed samples. In the present study, it was revealed that the two ultrawide-angle fundus imaging systems have high diagnostic sensitivity for retinal tears, and the Zeiss Clarus 500 system has higher diagnostic performance in temporal retinal tears, cataracts, and severe myopia.

In conclusion, Opel Panorama 200 and Zeiss Clarus 500 have the same sensitivity for analyzing retinal tears, Zeiss Clarus 500 has improved sensitivity under refractive opacity, and Clarus 500 has improved sensitivity for retinal tears in the temporal and far peripheral regions.

Acknowledgements

Not applicable.

Funding

Funding: No funding was received.

Availability of data and materials

The datasets used and/or analyzed during the present study are available from the corresponding author upon reasonable request.

Authors' contributions

YG and HY played a central role, contributing to various aspects of the research, from conceptualization to visualization, and taking the lead in writing. YG and CJG had a significant role, particularly in the methodology and supervision of the project. CJG and JW focused on data collection, data analysis, and contributed to editing and visualization. YG and HY confirm the authenticity of all the raw data. All authors read and approved the final version of the manuscript.

Ethics approval and consent to participate

The present study was approved (approval no. XPR-2021-0007) by the Xi'an BRIGHT Eye Hospital Ethics Committee (Xi'an, China). All patients underwent optometry, intraocular pressure, and visual acuity examinations, and they all voluntarily participated in the present study and signed informed consent forms.

Patient consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

References

1 

Ai S, Wang X, Ma M and Wang K: A method for correcting non-linear geometric distortion in ultra-wide-angle imaging system. Optik. 124:7014–7021. 2013.

2 

Nagiel A, Lalane RA, Sadda SR and Schwartz SD: Ultra-widefield fundus imaging: A review of clinical applications and future trends. Retina. 36:660–678. 2016.PubMed/NCBI View Article : Google Scholar

3 

Goh JKH, Cheung CY, Sim SS, Tan PC, Tan GSW and Wong TY: Retinal imaging techniques for diabetic retinopathy screening. J Diabetes Sci Technol. 10:282–294. 2016.PubMed/NCBI View Article : Google Scholar

4 

Panwar N, Huang P, Lee J, Keane PA, Chuan TS, Richhariya A, Teoh S, Lim TH and Agrawal R: e-Health, Fundus photography in the 21st century-a review of recent technological advances and their implications for worldwide healthcare. Telemed J E Health. 22:198–208. 2016.PubMed/NCBI View Article : Google Scholar

5 

Kaines A, Tsui I, Sarraf D and Schwartz S: The use of ultra wide field fluorescein angiography in evaluation and management of uveitis. Semin Ophthalmol. 24:pp19–24. 2009.PubMed/NCBI View Article : Google Scholar

6 

Oh BL, Park UC, Kim BH, Lee EK, Yoon CK, Choe HR, Yeon DY and Yu HG: Role of Ultra-widefield Imaging in the evaluation of Long-term change of highly myopic fundus. Acta Ophthalmol. 100:e977–e985. 2022.PubMed/NCBI View Article : Google Scholar

7 

Majumdar S and Tripathy K: Macular hole. In: StatPearls. StatPearls Publishing, Treasure Island, FL, 2023.

8 

Mohan R, Kohner EM, Aldington SJ, Nijhar I, Mohan V and Mather HM: Evaluation of a non-mydriatic camera in Indian and European diabetic patients. Br J Ophthalmol. 72:841–845. 1988.PubMed/NCBI View Article : Google Scholar

9 

Taylor R, Lovelock L, Tunbridge WM, Alberti KG, Brackenridge RG, Stephenson P and Young E: Comparison of non-mydriatic retinal photography with ophthalmoscopy in 2159 patients: Mobile retinal camera study. BMJ. 301:1243–1247. 1990.PubMed/NCBI View Article : Google Scholar

10 

Silva PS, Cavallerano JD, Sun JK, Soliman AZ, Aiello LM and Aiello LP: Peripheral lesions identified by mydriatic ultrawide field imaging: Distribution and potential impact on diabetic retinopathy severity. Ophthalmology. 120:2587–2595. 2013.PubMed/NCBI View Article : Google Scholar

11 

Oh K, Kang HM, Leem D, Lee H, Seo KY and Yoon S: Early detection of diabetic retinopathy based on deep learning and ultra-wide-field fundus images. Sci Rep. 11(1897)2021.PubMed/NCBI View Article : Google Scholar

12 

Fenner BJ, Wong RLM, Lam WC, Tan GSW and Cheung GCM: Advances in retinal imaging and applications in diabetic retinopathy screening: A review. Ophthalmol Ther. 7:333–346. 2018.PubMed/NCBI View Article : Google Scholar

13 

Toslak D, Chau F, Erol MK, Liu C, Chan RVP, Son T and Yao X: Trans-pars-planar illumination enables a 200° ultra-wide field pediatric fundus camera for easy examination of the retina. Biomed Opt Express. 11:68–76. 2019.PubMed/NCBI View Article : Google Scholar

14 

Khurram Butt D, Gurbaxani A and Kozak I: Ultra-wide-field fundus autofluorescence for the detection of inherited retinal disease in difficult-to-examine children. J Pediatr Ophthalmol Strabismus. 56:383–387. 2019.PubMed/NCBI View Article : Google Scholar

15 

Li S, Sun L, Liu C, Wang W, Huang S, Zhang T, Chen C, Wang Z, Cao L, Luo X, et al: Clinical features of ocular toxocariasis: A comparison between ultra-wide-field and conventional camera imaging. Eye (Lond). 35:2855–2863. 2021.PubMed/NCBI View Article : Google Scholar

16 

Hirano T, Imai A, Kasamatsu H, Kakihara S, Toriyama Y and Murata T: Assessment of diabetic retinopathy using two ultra-wide-field fundus imaging systems, the Clarus® and Optos systems. BMC Ophthalmol. 18(332)2018.PubMed/NCBI View Article : Google Scholar

17 

Ghasemi Falavarjani K, Wang K, Khadamy J and Sadda SR: Ultra-wide-field imaging in diabetic retinopathy; an overview. J Curr Ophthalmol. 28:57–60. 2016.PubMed/NCBI View Article : Google Scholar

18 

Price LD, Au S and Chong NV: Optomap ultrawide field imaging identifies additional retinal abnormalities in patients with diabetic retinopathy. Clin Ophthalmol. 9:527–531. 2015.PubMed/NCBI View Article : Google Scholar

19 

Kernt M, Pinter F, Hadi I, Hirneiss C, Haritoglou C, Kampik A, Ulbig MW and Neubauer AS: Diabetic retinopathy: Comparison of the diagnostic features of ultra-widefield scanning laser ophthalmoscopy Optomap with ETDRS 7-field fundus photography. Ophthalmologe. 108:117–123. 2011.PubMed/NCBI View Article : Google Scholar : (In German).

20 

Neubauer AS, Kernt M, Haritoglou C, Priglinger SG, Kampik A and Ulbig MW: Nonmydriatic screening for diabetic retinopathy by ultra-widefield scanning laser ophthalmoscopy (Optomap). Graefes Arch Clin Exp Ophthalmol. 246:229–235. 2008.PubMed/NCBI View Article : Google Scholar

21 

Kothari N, Pineles S, Sarraf D, Velez F, Heilweil G, Holland G, McCannel CA, Onclinx T, McCannel TA, Sadda SR, et al: Clinic-based ultra-wide field retinal imaging in a pediatric population. Int J Retina Vitreous. 5 (Suppl 1)(S21)2019.PubMed/NCBI View Article : Google Scholar

22 

Linz MO and Scott AW: Wide-field imaging of sickle retinopathy. Int J Retina Vitreous. 5 (Suppl 1)(S27)2019.PubMed/NCBI View Article : Google Scholar

23 

Latasiewicz M, Sala-Puigdollers A, Gonzalez-Ventosa A, Milla E and Adan Civera A: Multimodal retinal imaging of familial amyloid polyneuropathy. Ophthalmic Genet. 40:407–420. 2019.PubMed/NCBI View Article : Google Scholar

Related Articles

Journal Cover

February-2024
Volume 20 Issue 2

Print ISSN: 2049-9434
Online ISSN:2049-9442

Sign up for eToc alerts

Recommend to Library

Copy and paste a formatted citation
x
Spandidos Publications style
Guo Y, Yan H, Guo C and Wang J: Comparative analysis of two retinal fractures with ultrabroad‑angle fundus photography systems. Biomed Rep 20: 34, 2024
APA
Guo, Y., Yan, H., Guo, C., & Wang, J. (2024). Comparative analysis of two retinal fractures with ultrabroad‑angle fundus photography systems. Biomedical Reports, 20, 34. https://doi.org/10.3892/br.2024.1722
MLA
Guo, Y., Yan, H., Guo, C., Wang, J."Comparative analysis of two retinal fractures with ultrabroad‑angle fundus photography systems". Biomedical Reports 20.2 (2024): 34.
Chicago
Guo, Y., Yan, H., Guo, C., Wang, J."Comparative analysis of two retinal fractures with ultrabroad‑angle fundus photography systems". Biomedical Reports 20, no. 2 (2024): 34. https://doi.org/10.3892/br.2024.1722