Biomechanical comparison of bioresorbable F‑u‑HA/PLLA screw and tension band wire fixation in medial malleolar fractures

  • Authors:
    • Hiroaki Sakai
    • Naoya Takada
    • Gen Kuroyanagi
    • Takuya Usami
    • Yoshino Ueki
    • Hideki Murakami
  • View Affiliations

  • Published online on: March 31, 2021     https://doi.org/10.3892/wasj.2021.97
  • Article Number: 26
  • Copyright: © Sakai et al. This is an open access article distributed under the terms of Creative Commons Attribution License.

Metrics: Total Views: 0 (Spandidos Publications: | PMC Statistics: )
Total PDF Downloads: 0 (Spandidos Publications: | PMC Statistics: )


Abstract

The authors have recently reported that forged composites of raw particulate unsintered hydroxyapatite/poly‑L‑lactide (F‑u‑HA/PLLA) screws are useful for the treatment of medial malleolar fracture. However, the biomechanical properties of these screws have not yet been fully elucidated. The aim of the present study was to compare the biomechanical properties of F‑u‑HA/PLLA screws with those of tension band wiring (TBW) using medial malleolar fracture models obliquely oriented at 45˚ angles in synthetic tibial bones. Fractures were fixed using two 4.5‑mm‑diameter partial thread F‑u‑HA/PLLA screws, two 4.5‑mm‑diameter full thread F‑u‑HA/PLLA screws, or TBW. The fracture models were tested parallel to the distal joint surface and parallel to the fracture line. The compression load at 1, 2, 3, 4 and 5 mm of fracture displacement was assessed. The compression load parallel to the distal joint surface at 3, 4 and 5 mm of fracture displacement was significantly higher in the partial thread screw group than in the TBW group (P<0.01). Additionally, the compression load parallel to the distal joint surface at 4 and 5 mm was significantly higher in the full thread screw group than in the TBW group (P<0.001). However, the TBW group exhibited a higher compression load at 1 mm compared to the full thread group (P<0.05). As regards the compression load parallel to the fracture line, the compression load at 3, 4 and 5 mm of fracture displacement was markedly higher in the partial thread screw group than in the TBW group (P<0.01); likewise, the compression load at 4 and 5 mm of displacement was significantly higher in the full thread screw group than in the TBW group (P<0.001). On the whole, these results strongly suggest that F‑u‑HA/PLLA screw fixation, particularly partial thread screw fixation, has good biomechanical strength and provides adequate stability for the treatment of medial malleolar fractures.

Introduction

Medial malleolar fractures are very common and are associated with ankle injury (1). Since the medial malleolus plays a pivotal role in the stability and function of the ankle joint, anatomical reduction by surgery is generally recommended for displaced medial malleolar fractures (2). Reportedly, open reduction and internal fixation restores articular congruency and prevents post-traumatic arthritis (2). A variety of techniques for medial malleolar fracture repair employ metallic implants, including metallic screws and tension band wiring (TBW) by Kirshner wires (K-wires) and stainless-steel wire (3-5). Although a number of studies have described good clinical outcomes following surgical treatment, post-operative complications, such as residual pain, skin irritation and the backing out of implants caused by metallic implants are problematic (3-5). Moreover, in some cases, the patient requires the removal of the metallic implant due to these post-operative complications (6).

The implant bioactivity, bioabsorbability and high mechanical strength of forged composites of raw particulate unsintered hydroxyapatite/poly-L-lactide (F-u-HA/PLLA) are similar to those of human cortical bone (7). The authors have previously demonstrated the utility of F-u-HA/PLLA screws and pins for the treatment of tibial plateau fractures and distal humeral fractures (8,9). Furthermore, F-u-HA/PLLA screws have been proven effective for the treatment of medial malleolar fractures, demonstrating good clinical results without any post-operative complications, such as deep infections, late aseptic tissue response, or foreign body reactions (10). Additionally, none of the patients required the removal of these screws (10). However, to the best of our knowledge, there are no reports available to date of mechanical comparisons between F-u-HA/PLLA screws and TBW in medial malleolar fractures. Thus, the aim of the present study was to compare the biomechanical properties of partial or full thread F-u-HA/PLLA screws with TBW in a medial malleolar fracture model using Synbone® AG.

Materials and methods

Fracture models and grouping

A total of 42 fourth generation composite synthetic left tibial bones (Synbone® AG) were used in the present study. The tibial bone was cut 17 cm from the distal articular surface. Medial malleolar fractures obliquely oriented at 45˚ angles to the longitudinal axis of the tibial bone were created using a saw.

The 42 fracture models were divided into three groups as follows: a TBW group, a partial thread screw group and a full thread screw group (n=14 per group). The TBW group (n=14) underwent traditional TBW using two 1.6-mm-diameter K-wires and a 0.9-mm-diameter stainless-steel wire (Mizuho Co., Ltd.) (Fig. 1A). The stainless-steel wire was passed through the tibia on the proximal side of the fracture line, and then tightened in a figure-eight pattern. The tension of the stainless-steel wire was kept constant at 30 N by a scale device. The partial thread screw group (n=14) was fixed with two partial thread 4.5x45 mm F-u-HA/PLLA screws (Osteotrans Puls®, Teijin Medical Technologies Co., Ltd.), which were fixed mono-cortically (Fig. 1B). On the other hand, the full thread screw group (n=14) was fixed with two full thread 4.5x70 mm F-u-HA/PLLA screws that were fixed bi-cortically (Fig. 1C). The universal testing machine (AG-20kNXD, Shimadzu Corporation) tested the tibial bone models under two compression loading conditions. The load was applied parallel to the distal tibial joint surface and parallel to the fracture line at 10 mm/min, respectively (Fig. 2). The direction of the load was simulated as the force received from the talus during walking and the force pushed up by the talus during a fracture. Compression loads at 1, 2, 3, 4 and 5 mm of fracture displacement were analyzed.

Figure 1

(A) The tension band wiring group was fixed using two 1.6-mm-diameter K-wires and a 0.9-mm-diameter metallic soft wire. (B) The partial thread screw group was fixated with two partial thread 4.5x45 mm unsintered hydroxyapatite particles/poly-L-lactide (F-u-HA/PLLA) screws. (C) The full thread screw group was fixated with two full thread 4.5x70 mm F-u-HA/PLLA screws.

Figure 2

Image showing the tension loading mechanism. (A) Compression load parallel to the distal joint surface (B) Compression load parallel to the fracture line.

Statistical analysis

Statistical analyses were performed with SPSS version 23 statistical software (IBM Co.). Data were analyzed using one-way analysis of variance (ANOVA) with Tukey's honestly significant difference (HSD) post hoc test for multiple comparisons. A P-value <0.05 was considered to indicate a statistically significant difference. All data are presented as the mean ± standard error of the mean (SEM) of seven determinations from independent model preparations.

Results

The compression load parallel to the distal joint surface at 3, 4 and 5 mm of fracture displacement in the partial thread screw group was significantly higher than that in the TBW group (3 mm: TBW, 84±8 N; partial thread, 154±50 N; 4 mm: TBW, 93±9 N; partial thread, 196±32 N; 5 mm: TBW, 99±10 N; partial thread, 221±26 N; P<0.01). Similar to the partial thread screw, the compression load parallel to the distal joint surface at 4 and 5 mm of fracture displacement in the full thread screw group was significantly higher than that in the TBW group (full thread: 4 mm, 180±19 N; 5 mm, 246±19 N; P<0.001). However, the compression load parallel to the distal joint surface at 1 mm of fracture displacement in the full thread screw group was significantly lower than that in the TBW group (1 mm: TBW, 47±6 N; full thread, 25±5 N; P<0.05) (Fig. 3).

Figure 3

Compression load parallel to the distal joint surface was calculated under the fracture displacement. Each value represents the mean ± standard error of the mean (SEM) of seven determinations from independent model preparations. *P<0.01, **P<0.001 and ***P<0.001 compared to the value of the TBW group. TBW, tension band wiring.

As regards the compression load parallel to the fracture line, the partial thread screw group exhibited a significantly higher compression load than the TBW group between 3 and 5 mm (3 mm: TBW, 187±18 N; partial thread, 245±28 N; 4 mm: TBW, 225±18 N; partial thread, 310±30 N; 5 mm: TBW, 262±20 N; partial thread, 378±25 N; P<0.01). The compression load parallel to the fracture line at 4 and 5 mm of fracture displacement in the full thread screw group was significantly higher than that in the TBW group (full thread: 4 mm, 295±32 N; 5 mm, 378±34 N; P<0.001). There were no significant differences in the compression load parallel to the fracture line at 1 and 2 mm of fracture displacement between the TBW group and the partial or full thread screw groups (Fig. 4). No breakage of the F-u-HA/PLLA screws or TBW was observed when the gap between fractures reached 5 mm.

Figure 4

Compression load parallel to the fracture line was calculated under the fracture displacement. Each value represents the mean ± standard error of the mean (SEM) of seven determinations from independent model preparations. **P<0.01 and ***P<0.001 compared to the value of the TBW group. TBW, tension band wiring.

Discussion

The present study demonstrated that partial thread F-u-HA/PLLA screws are stronger than TBW in terms of the fixation of medial malleolar fractures. On the other hand, full thread F-u-HA/PLLA screws also exhibited stronger compression loads with a displacement of >4 mm under both conditions, parallel to the distal joint surface and parallel to the fracture line, compared to TBW. However, when the displacement was 1 mm parallel to the distal joint surface, the compression load tolerance of full thread F-u-HA/PLLA screws was lower than that of TBW. TBW is usually firmly fixed by crimping even when the displacement of the fractured part is small. On the other hand, following a displacement of 2 mm, since the tension of the stainless steel wire did not function and the K-wire itself was easily pulled out, it was probable that the value of the compression load remained low. As for the full thread screws, as the full thread screw is bi-cortically fixed and not backed out, it is likely that the tensile strength of the screw greatly affected the load value.

It has been reported that full thread metal screws have a higher strength than that of partial thread metal screws for the fixation of medial malleolar fractures (3,11). Fowler et al (11) reported that bi-cortical screw fixation by full thread screws is superior to uni-cortical screw fixation by partial thread screws in terms of transverse (vs. axial) compression loads at 2 mm of displacement using tibial sawbone fracture models. Downey et al (12) also reported that full thread screws exhibited a higher initial construct stiffness and catastrophic failure strength compared to partial thread screws following shear axial loading using sawbone blocks. Thus, when using the metal implant, it is likely that bi-cortical screw fixation by full thread screws has an advantage over partial thread screws. However, the present study demonstrated that the compression load of the partial thread F-u-HA/PLLA screw group was stronger in displacements both parallel to the distal joint surface and parallel to the fracture line than that of the full thread F-u-HA/PLLA screw group. Based on these findings, the mechanical properties of F-u-HA/PLLA screws and metallic screws differ in medial malleolar fractures. Thus, further studies on screw suitability in the treatment of medial malleolar fractures are required. The screws used in the present study are made from composites of uncalcined and unsintered HA particles and PLLA. The material properties of u-HA/PLLA composite are summarized in Table I.

Table I

Material properties of the u-HA/PLLA composite.

Table I

Material properties of the u-HA/PLLA composite.

Property Value Unit
Viscosity average molecular weight 210 kDa
Bending strength 269.2 (±2.5) MPa
Bending modulus 7.6 (±0.1) GPa
Tensile strength 121.5 (±6.2) MPa
Tensile modulus 2.3 (±0.2) GPa
Compressive strength 106.7 (±3.7) MPa
Shear strength 126.4 (±3.3) MPa
Torsional strength (Ø3.2 mm) 64.7 (±1.0) N·cm

TBW fixation is one of the most common treatments recommended by the Arbeitsgemeinschaft für Osteosynthesefragen (AO) group for medial malleolar fractures (13). It has been shown that fixation by TBW is stronger than fixation by uni-cortical partial thread metallic screws using a cadaver model (5). However, due to post-operative complications caused by metallic wire and pins, implant removal is often required (6). Recently, the AO group recommended fixation using two parallel partial threads cancellous screws for the treatment of medial malleolar fractures (14). Previously, the authors demonstrated that F-u-HA/PLLA partial thread screws were effective for the treatment of medial malleolar fractures without any post-operative complications (10). Additionally, the results of the present study demonstrated that F-u-HA/PLLA partial thread screws were stronger than TBW. Taking these findings into account, F-u-HA/PLLA partial thread screws may be an alternative option for the treatment of medial malleolar fractures.

F-u-HA/PLLA implants have several advantages, including bioactivity, bioresorbability and a high mechanical strength (7). These implants also have good biocompatibility and osteoconductivity, which stimulate direct bone-bonding and new bone formation (15,16). Thus, F-u-HA/PLLA implants are currently utilized in several clinical fields, such as orthopedic, maxillofacial and trauma surgeries (17-19). Compared to other bioresorbable PLLA implants, the F-u-HA/PLLA implant has higher initial mechanical strength than human cortical bone, which has a strength of 200 MPa (7). Shikinami and Okuno (7) reported that the F-u-HA/PLLA material had an initial bending strength of 270 MPa, which is much higher than that of natural cortical bone. In a previous study by the authors, the breakage of F-u-HA/PLLA screws was not observed following the treatment of medial malleolar fractures (10). Although there is no doubt that F-u-HA/PLLA screws are generally not applicable in the treatment of load-bearing joints, F-u-HA/PLLA screws seem to be strong enough for use in this area.

There are several limitations to the present study. First, Synbone® AG was used in the present study. Synbone® AG is a synthetic sawbone and is known to have biomechanical properties similar to those of human bone (20). However, Synbone® AG mimics normal healthy bone structure; thus, the present results may not be applicable to patients with osteoporosis. Second, the mechanical test was performed under only two load conditions. In a clinical setting, the fracture site is subject to a variety of forces from various directions through the surrounding tissues. Thus, further research with other load conditions will be required to elucidate the detailed mechanical properties of the F-u-HA/PLLA screws. Finally, medial malleolus fracture with fracture line located at 45˚ to the joint space is not the most common in clinical practice. Generally, medial malleolus fractures can be distinguished from fracture types, such as parallel or perpendicular location to the joint line. Forces acting on different types of malleolar fractures may have different influence on the screws. Therefore, it seems likely that the results presented herein may not be directly applicable to clinical practice. Further investigations are required to clarify the effects of different fracture patterns on the screws.

In conclusion, the present study demonstrates that F-u-HA/PLLA screw fixation has a good biomechanical strength and provides adequate stability for the treatment of medial malleolar fractures. The F-u-HA/PLLA screw may prove to be a useful option for the treatment of medial malleolar fractures.

Acknowledgements

The F-u-HA/PLLA screws, Synbone® AG, and the universal testing machine were kindly provided by Takiron Co. Ltd.

Funding

The present study was supported in part by a Grant‑in‑Aid for Scientific Research (19K18471) from the Ministry of Education, Culture, Sports, Science and Technology of Japan.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Authors' contributions

HS and NT, YU and HM designed the study. HS, NT and TU performed the experiments. HS, NT and GK analyzed the data. HS, NT and GK drafted the manuscript. NT, YU and HM critically reviewed the manuscript. HS and GK confirm the authenticity of all the raw data. All authors have read and approved the final manuscript.

Ethics approval and consent to participate

No human tissues were used in the present study; only synthetic bone tissue was used.

Patient consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

References

1 

Lokerman RD, Smeeing DPJ, Hietbrink F, van Heijl M and Houwert RM: Treatment of a Scientifically Neglected Ankle Injury: The Isolated Medial Malleolar Fracture. A Systematic Review. J Foot Ankle Surg. 58:959–968. 2019.PubMed/NCBI View Article : Google Scholar

2 

Carter TH, Duckworth AD and White TO: Medial malleolar fractures: Current treatment concepts. Bone Joint J. 101B:512–521. 2019.PubMed/NCBI View Article : Google Scholar

3 

Yammine K, Karam K and Assi C: Cortical versus cancellous screws in treating medial malleolar fractures: A systematic review of comparative clinical and biomechanical studies. Foot. 40:81–86. 2019.PubMed/NCBI View Article : Google Scholar

4 

Barnes H, Cannada LK and Watson JT: A clinical evaluation of alternative fixation techniques for medial malleolus fractures. Injury. 45:1365–1367. 2014.PubMed/NCBI View Article : Google Scholar

5 

Ebraheim NA, Ludwig T, Weston JT, Carroll T and Liu J: Comparison of surgical techniques of 111 medial malleolar fractures classified by fracture geometry. Foot Ankle Int. 35:471–477. 2014.PubMed/NCBI View Article : Google Scholar

6 

Brown OL, Dirschl DR and Obremskey WT: Incidence of hardware-related pain and its effect on functional outcomes after open reduction and internal fixation of ankle fractures. J Orthop Trauma. 15:271–274. 2001.PubMed/NCBI View Article : Google Scholar

7 

Shikinami Y and Okuno M: Bioresorbable devices made of forged composites of hydroxyapatite (HA) particles and poly-L-lactide (PLLA): Part I. Basic characteristics. Biomaterials. 20:859–877. 1999.PubMed/NCBI View Article : Google Scholar

8 

Kuroyanagi G, Yoshihara H, Yamamoto N, Suzuki H, Yamada K, Yoshida Y, Otsuka T and Takada N: Treatment of Lateral Tibial Condylar Fractures Using Bioactive, Bioresorbable Forged Composites of Raw Particulate Unsintered Hydroxyapatite/Poly-L-Lactide Screws. Orthopedics. 41:e365–e368. 2018.PubMed/NCBI View Article : Google Scholar

9 

Takada N, Suzuki H, Yamada K and Otsuka T: Treatment of Ankle Fractures Using New High-Strength, Bioactive, Bioresorbable Forged Composites of Unsintered Hydroxyapatite/poly-L-lactide (F-u-HA/PLLA) Implants. J Orthop. 7(e5)2010.

10 

Takada N, Otsuka T, Suzuki H and Yamada K: Pediatric displaced fractures of the lateral condyle of the humerus treated using high strength, bioactive, bioresorbable F-u-HA/PLLA pins: A case report of 8 patients with at least 3 years of follow-up. J Orthop Trauma. 27:281–284. 2013.PubMed/NCBI View Article : Google Scholar

11 

Fowler TT, Pugh KJ, Litsky AS, Taylor BC and French BG: Medial malleolar fractures: A biomechanical study of fixation techniques. Orthopedics. 34:e349–e355. 2011.PubMed/NCBI View Article : Google Scholar

12 

Downey MW, Kosmopoulos V and Carpenter BB: Fully Threaded Versus Partially Threaded Screws: Determining Shear in Cancellous Bone Fixation. J Foot Ankle Surg. 54:1021–1024. 2015.PubMed/NCBI View Article : Google Scholar

13 

Meinberg EG, Agel J, Roberts CS, Karam MD and Kellam JF: Fracture and dislocation classification compendium-2018. J Orthop Trauma. 32 (Suppl 1):S1–S170. 2018.PubMed/NCBI View Article : Google Scholar

14 

Parker L, Garlick N, McCarthy I, Grechenig S, Grechenig W and Smitham P: Screw fixation of medial malleolar fractures: A cadaveric biomechanical study challenging the current AO philosophy. Bone Joint J. 95-B:1662–1666. 2013.PubMed/NCBI View Article : Google Scholar

15 

Furukawa T, Matsusue Y, Yasunaga T, Nakagawa Y, Okada Y, Shikinami Y, Okuno M and Nakamura T: Histomorphometric study on high-strength hydroxyapatite/poly(L-lactide) composite rods for internal fixation of bone fractures. J Biomed Mater Res. 50:410–419. 2000.PubMed/NCBI View Article : Google Scholar

16 

Hasegawa S, Ishii S, Tamura J, Furukawa T, Neo M, Matsusue Y, Shikinami Y, Okuno M and Nakamura T: A 5-7 year in vivo study of high-strength hydroxyapatite/poly(L-lactide) composite rods for the internal fixation of bone fractures. Biomaterials. 27:1327–1332. 2006.PubMed/NCBI View Article : Google Scholar

17 

Lind M, Feller J and Webster KE: Tibial bone tunnel widening is reduced by polylactate/hydroxyapatite interference screws compared to metal screws after ACL reconstruction with hamstring grafts. Knee. 16:447–451. 2009.PubMed/NCBI View Article : Google Scholar

18 

Li ZH, Yu AX, Guo XP, Qi BW, Zhou M and Wang WY: Absorbable implants versus metal implants for the treatment of ankle fractures: A meta-analysis. Exp Ther Med. 5:1531–1537. 2013.PubMed/NCBI View Article : Google Scholar

19 

Landes C, Ballon A, Ghanaati S, Tran A and Sader R: Treatment of malar and midfacial fractures with osteoconductive forged unsintered hydroxyapatite and poly-L-lactide composite internal fixation devices. J Oral Maxillofac Surg. 72:1328–1338. 2014.PubMed/NCBI View Article : Google Scholar

20 

Culemann U, Holstein JH, Köhler D, Tzioupis CC, Pizanis A, Tosounidis G, Burkhardt M and Pohlemann T: Different stabilisation techniques for typical acetabular fractures in the elderly - a biomechanical assessment. Injury. 41:405–410. 2010.PubMed/NCBI View Article : Google Scholar

Related Articles

Journal Cover

May-June 2021
Volume 3 Issue 3

Print ISSN: 2632-2900
Online ISSN:2632-2919

Sign up for eToc alerts

Recommend to Library

Copy and paste a formatted citation
x
Spandidos Publications style
Sakai H, Takada N, Kuroyanagi G, Usami T, Ueki Y and Murakami H: Biomechanical comparison of bioresorbable F‑u‑HA/PLLA screw and tension band wire fixation in medial malleolar fractures. World Acad Sci J 3: 26, 2021
APA
Sakai, H., Takada, N., Kuroyanagi, G., Usami, T., Ueki, Y., & Murakami, H. (2021). Biomechanical comparison of bioresorbable F‑u‑HA/PLLA screw and tension band wire fixation in medial malleolar fractures. World Academy of Sciences Journal, 3, 26. https://doi.org/10.3892/wasj.2021.97
MLA
Sakai, H., Takada, N., Kuroyanagi, G., Usami, T., Ueki, Y., Murakami, H."Biomechanical comparison of bioresorbable F‑u‑HA/PLLA screw and tension band wire fixation in medial malleolar fractures". World Academy of Sciences Journal 3.3 (2021): 26.
Chicago
Sakai, H., Takada, N., Kuroyanagi, G., Usami, T., Ueki, Y., Murakami, H."Biomechanical comparison of bioresorbable F‑u‑HA/PLLA screw and tension band wire fixation in medial malleolar fractures". World Academy of Sciences Journal 3, no. 3 (2021): 26. https://doi.org/10.3892/wasj.2021.97