Open Access

Clinical effect of a temporary pacemaker on an electrical storm induced by severe acute aconitine poisoning

  • Authors:
    • Feng Cheng
    • Li Hao
    • Ming-Wei Liu
  • View Affiliations

  • Published online on: January 31, 2022     https://doi.org/10.3892/wasj.2022.142
  • Article Number: 7
  • Copyright: © Cheng et al. This is an open access article distributed under the terms of Creative Commons Attribution License.

Metrics: Total Views: 0 (Spandidos Publications: | PMC Statistics: )
Total PDF Downloads: 0 (Spandidos Publications: | PMC Statistics: )


Abstract

In China, medicinal liquor made from the aconitine root soaked in wine is poisoned. At present, reliable detoxification drugs are not available. The present study used a temporary pacemaker to treat 12 patients who required treatment due to an electrical storm caused by aconitine poisoning between June, 2017 and April, 2021. The results revealed that temporary pacing was successful in all 12 patients. The electrode placement rate was 100%, and the effective cardiac pacing rate was 100%. The blood pressure of the patients gradually returned to normal levels. Ventricular tachycardia and high‑grade atrioventricular block were terminated. As regards atrial premature beats and ventricular premature beats, the sinus rhythm was restored after continuous amiodarone pumping. Simultaneously, fluid supplementation and control supportive treatment were administered. The patients were hospitalized for 4‑6 days. The patients' symptoms disappeared, their vital signs were stable and their electrocardiogram (ECG) exhibited normal findings. The 12 patients were followed‑up for 1 week following discharge and did not have any discomfort; the ECG revealed no abnormalities. On the whole, the findings of the present study suggest that temporary cardiac pacing may be an effective treatment for malignant arrhythmias caused by severe acute aconitine poisoning. However, the present study only involved a very small number of cases. In the future, further studies are required with larger sample, and further randomized, controlled multi‑center studies are also warranted to these findings.

Introduction

Aconitine exerts pharmacological effects, such as dispelling wind and dampness, and easing menstrual cramps and pain (1). In China, it is considered common practice to soak Radix aconitiagrestis and Radix aconiti sinensisin wine or to add Kusnezoff monkshood root slices when cooking food. However, the majority of plants of the genus Aconitum are poisonous. Aconitine is the main toxic component of Aconitum plants. It exerts effects mainly on the heart and nervous system, causing a variety of arrhythmias and leading to death (1). Its therapeutic dose is close to the poisonous or lethal dose, and it is a typical representative of poisonous plants with medicinal value (2). The oral administration of 0.2 mg aconitine can be poisonous, and a dose of 3-5 mg can cause death. During use, aconitine can easily cause poisoning and death due to individual differences in drug tolerance, improper usage or misuse and deliberate poisoning (3). The cause of death is mostly malignant arrhythmia, and the hospital mortality rate in China is 5.5% (3). In China, medicinal liquor made from aconitine root soaked in wine is mostly poisoned.

At present, reliable detoxification drugs are not available. In addition to inducing vomiting, gastric lavage and catharsis, the key to treating aconitine poisoning is to correct arrhythmia in a timely and effective manner, stabilize myocardial cells, and maintain the stability of vital signs (4). The majority of patients are treated with atropine, lidocaine, or amiodarone, which can effectively control arrhythmia and cure patients. However, some severely ill patients have severe arrhythmia, such as an electrical storm, and thus do not survive due to ineffective treatment with antiarrhythmic drugs. Temporary pacemakers enable pacing and biphasic defibrillation, which can pace the heart in a short period of time, and also defibrillate, restore autonomous heart rhythm, prevent malignant arrhythmia, and ensure blood supply to vital organs (5). To the best of our knowledge, no study to date has used temporary pacemakers for the treatment of malignant arrhythmias induced by aconitine poisoning. Therefore, the present study used a temporary pacemaker to treat 12 patients who were admitted to the Department of Emergency Medicine of Yan'an Hospital in Kunming from June, 2017 to April, 2021 due to an electrical storm caused by aconitine poisoning, and all were cured.

Patients and methods

Patient information

From June, 2017 to April, 2021, 12 patients were admitted to the Department of Emergency Medicine of Yan'an Hospital in Kunming due to aconitine poisoning. Among them, 8 patients were males, and 4 patients were females. The patients' age ranged from 33 to 65 years, with an average age of 45 years. Poisoning was induced in 2 patients due to the ingestion of herbal black powder, and in 10 patients due to the consumption of medicated wine. Symptoms occurred after 10-30 min, and they all visited the Department of Emergency Medicine within 1-2 h following the onset of symptoms. The present study was approved by the First Affiliated Hospital of Kunming Medical University Research Ethics Boards. Written informed consent for participation in the study was obtained from the parent or relative of the participants.

Inclusion criteria

The inclusion criteria were as follows: i) All the 12 patients met the diagnostic criteria for ventricular tachycardia electrical storm: Patients who suffered from two or more spontaneous ventricular tachycardias or ventricular fibrillation within 24 h, sharp and severe disturbance of cardiac electrical activity, rapid ventricular tachycardia, and ventricular fibrillation that can occur repeatedly requiring repeated electrical cardioversion; ii)treatment with atropine, lidocaine, or amiodarone was ineffective; iii) electrical cardioversion therapy was ineffective; iv) suffered from combined high-grade atrioventricular block; v) suffered from QT syndrome with torsades de pointes; vi) patients and relatives agreed to the use of temporary pacemakers.

Exclusion criteria

The exclusion criteria were as follows: i) Patients in which a ventricular tachycardia electrical storm was caused by non-aconitine poisoning; ii) treatment with atropine, lidocaine, or amiodarone was effective; iii) electrical cardioversion was effective; iv) patients and relatives did not agree to the use of temporary pacemakers.

Clinical manifestations

All 12 cases exhibited symptoms, such as numbness in the mouth, tongue, face and limbs, and other symptoms including nausea, vomiting, dizziness, sweating, chest tightness and palpitations. Among the patients, there were two cases of convulsions with unconsciousness during medical treatment, six cases of syncope, two cases of convulsions with facial cyanosis, and two cases of whole-body numbness and a ‘feeling of dying’ (Table I).

Table I

Clinical data of the 12 patients before treatment with the temporary pacemaker.

Table I

Clinical data of the 12 patients before treatment with the temporary pacemaker.

Patient no.SexAge, yearsType of poison ingestedClinical manifestations and electrocardiogramTreatmentEffect
1Male41Traditional Chinese medicine (Radix aconiti agrestis)Chest tightness, vomiting, perioral numbness, convulsions, etc.; electrocardiogram: Multi-source premature ventricular contractions, short bursts of ventricular tachycardia, ventricular fibrillation; blood pressure: 78/49 mmHgAmiodarone, dopamine, lidocaine, a tropine, bedside gastric lavage, electrical defibrillation, symptomatic treatmentTreatment ineffective
2Male51Medicinal wine (Radix aconitiagrestis + wine)Palpitation, sweating, dizziness, syncope, etc.; ECG chart: Polymorphic ventricular tachycardia, ventricular premature contraction; blood pressure: 89/57 mmHgAmiodarone, lidocaine, atropine, dopamine, electrical defibrillation, bedside gastric lavage, symptomatic treatmentTreatment ineffective
3Female35Medicinal wine (Radix aconitiagrestis + wine)Chest tightness, general numbness, convulsions with facial cyanosis; ECG: Ventricular fibrillation, supraventricular tachycardia, frequent ventricular premature contractions; blood pressure: 61/38 mmHgElectric defibrillation, lidocaine, atropine, dopamine, potassium and magnesium supplements, symptomatic treatmentTreatment ineffective
4Female48Medicinal wine (Radix aconitiagrestis + wine)Chest tightness, palpitations, sweating, syncope, etc.; ECG: Torsades de pointes, ventricular tachycardia; blood pressure: 91/52 mmHgElectric defibrillation, lidocaine, amiodarone, dopamine, potassium and magnesium supplements, symptomatic treatmentTreatment ineffective
5Female44Traditional Chinese medicine (Radix aconitiagrestis)Whole-body numbness and a ‘feeling of dying’, sweating, etc.; ECG: Multi-source ventricular premature contractions, torsades de pointes ventricular tachycardia; blood pressure: 95/61 mmHgElectric defibrillation, lidocaine, amiodarone, dopamine, potassium and magnesium supplements, symptomatic treatmentTreatment ineffective
6Male47Medicinal wine (Radix aconitiagrestis + wine)Whole-body numbness and a ‘feeling of dying’, sweating, etc.; ECG: Multi-source ventricular premature beats, ventricular tachycardia, ventricular fibrillation; blood pressure 95/61 mmHgElectric defibrillation, lidocaine, amiodarone, dopamine, potassium and magnesium supplements, symptomatic treatmentTreatment ineffective
7Male55Medicinal wine (Radix aconitiagrestis + wine)Palpitation, whole-body numbness and a ‘feeling of dying’, vomiting, etc.; ECG: Multi-source premature ventricular contractions, polymorphic ventricular tachycardia; blood pressure: 66/40 mmHgLidocaine, amiodarone, dopamine, potassium and magnesium supplements, symptomatic treatmentTreatment ineffective
8Male53Medicinal wine (Radix aconitiagrestis + wine)Perioral, tongue, facial numbness, vomiting, syncope with convulsions; ECG: Polymorphic ventricular tachycardia, short bursts of ventricular tachycardia; blood pressure: 71/38 mmHgElectric defibrillation, lidocaine, dopamine, potassium and magnesium supplements, symptomatic treatmentTreatment ineffective
9Female48Medicinal wine (Radix aconitiagrestis + wine)Dizziness, vomiting, chest tightness, general numbness, etc.; ECG: Polymorphic ventricular tachycardia, short bursts of ventricular tachycardia, ventricular fibrillation; blood pressure: 99/49 mmHgElectric defibrillation, amiodarone, lidocaine, bedside gastric lavageTreatment ineffective
10Male51Medicinal wine (Radix aconitiagrestis + wine)Chest tightness, palpitations, sweating, dizziness, vomiting, etc.; ECG: Polymorphic ventricular tachycardia, polygenic premature ventricular contractions; blood pressure: 82/53 mmHgLidocaine, defibrillation, dopamine, potassium and magnesiumTreatment ineffective
11Male38Medicinal wine (Radix aconitiagrestis + wine)Dizziness, vomiting, chest tightness, general numbness, etc.; ECG: High atrioventricular block, short bursts of ventricular tachycardia, multi-source ventricular premature contractions; blood pressure: 67/54 mmHgAmiodarone, atropine, lidocaine, bedside gastric lavageTreatment ineffective
12Male61Medicinal wine (Radix aconitiagrestis + wine)Chest tightness, palpitations, sweating, syncope, vomiting, etc.; ECG: Polymorphic ventricular tachycardia, QT syndrome with torsades de pointes; blood pressure: 91/53 mmHgAtropine, lidocaine, bedside gastric lavageTreatment ineffective

[i] ECG, electrocardiogram.

Electrocardiogram (ECG)

The first ECG or ECG monitoring revealed five cases of multi-source ventricular premature beats, six cases of polymorphic ventricular tachycardia, four cases of ventricular fibrillation, and three cases of torsades de pointes ventricular tachycardia. All 12 cases had varying degrees of hypotension (Table I). A typical ECG is illustrated in Fig. 1.

Treatment method

The 12 patients were immediately administered a thorough gastric lavage, activated charcoal plus 250 ml of mannitol oral catharsis, continuous ECG monitoring, oxygen inhalation, routine fluid replacement, intravenous supplementation of potassium and magnesium polarized fluids, correction of acidosis and electrolyte disorders, and 102 mg of intravenous atropine. Furosemide (Shaanxi Jingxi Pharmaceutical Co., Ltd.; 20 mg) is an intravenous diuretic. Lidocaine (Zhejiang Chengxin Pharmaceutical Co., Ltd.; 100 mg) was injected intravenously every 15-30 min. Amiodarone (Shandong Fangming Pharmaceutical Group Co., Ltd.) was used when lidocaine (Zhejiang Chengxin Pharmaceutical Co., Ltd.) was ineffective. All patients were administered a slow intravenous bolus of 150 mg of amiodarone (Shandong Fangming Pharmaceutical Group Co., Ltd.), followed by continuous micropump pumping of 1 mg/min amiodarone (Shandong Fangming Pharmaceutical Group Co., Ltd.). The aforementioned ‘rescue’ methods were not effectivein 10 patients, who manifested a loss of consciousness, shock, ventricular fibrillation and Adams-Stokes syndrome. The patients were immediately administered electric shock defibrillation, chest compressions and emergency bedside temporary cardiac pacing for rescue.

Equipment

The equipment used included a Medtronic Temporary Pacemaker T10 (Shanghai Zanzhuo Medical Instrument Co., Ltd.), no. 18 thin-walled puncture needle and J-shaped guide wire (Shenzhen Yixinda Medical New Technology Co., Ltd.), 6F bipolar electrode catheter (Beijing Wandong Medical Equipment Co., Ltd.), 6F venous dilator and venous sheath (Beijing Wandong Medical Equipment Co., Ltd.), a photoelectric electrocardiograph and a Minadray 9000 ECG monitor (Nanjing Baden Medical Co., Ltd.).

Operation method

For the procedure, the patient was placed on his/her back and monitored using an ECG. Following local infiltration anesthesia, the left subclavian vein was punctured by Seldinger's method, and the pacing electrode was successfully inserted. Bedside ECG monitoring was used for temporary cardiac pacing. After the puncture was successful, the outer sheath was rapidly fed into the 6F electrode catheter with a slightly curved tip end, and the electrode catheter was fed 20-25 cm. A temporary pacemaker was connected. The pacemaker voltage and sensitivity were adjusted to 3 and 0.8 mV, respectively. The pacing frequency was 60 or 10 beats/min higher than the patient's own frequency. Once the main wave in leads II, III and aVF appeared downward, and lead V1 presented a pacing pattern with left bundle branch block wide malformation, the ECG was traced. The pacing threshold <1.0 V was determined, and the lead was fixed and placed locally. Finally, the operation was complete.

Results

Effect of temporary pacemaker on wind burst induced by severe acute aconitine poisoning

Temporary pacing was successful in all 12 patients. The electrode placement rate was 100%, and the effective cardiac pacing rate was 100%. The patients' blood pressure gradually returned to normal. Ventricular tachycardia and high-grade atrioventricular block were terminated. For atrial premature beats and ventricular premature beats, the sinus rhythm was restored following continuous amiodarone pumping (Fig. 2). At the same time, fluid supplementation and control supportive treatment were administered. The patients were hospitalized for 4-6 days. The patients' symptoms disappeared, their vital signs were stable and their ECG exhibited normal findings. No abnormalities in the ECG were observed for 2 days, and the patients were discharged for follow-up.

Side-effects of the temporary pacemaker

In total, 1 patient experienced pacing dysfunction caused by temporary pacemaker displacement. Pacemaker malfunction was corrected after adjusting the position of the pacemaker.

Patient follow-up

The 12 patients were followed-up for 1 week following discharge without any discomfort, and the ECG did not reveal any abnormalities (Fig. 3).

Discussion

Aconitine mainly damages the circulatory and central nervous systems. The effects of aconitine poisoning on the heart excite the vagus nerve, causing the postganglionic fibers to release a large amount of acetylcholine, thereby reducing the autonomy and conductivity of the sinus node. Aconitine also has a direct effect on the myocardium, which renders the excitatory conduction and refractory period of each part of the myocardium inconsistent. Moreover, repolarization is not synchronized to form re-entry, resulting in severe ventricular arrhythmia. Ventricular tachycardia is the most common and likely cause of death. Aconitine can directly act on ventricular muscles, producing high-frequency ectopic rhythms and causing ventricular fibrillation, cardiogenic shock and Adams-Stokes syndrome (6). Therefore, malignant arrhythmia induced by acute aconitine poisoning can be life-threatening if not corrected in a timely manner.

Temporary cardiac pacemakers have biphasic defibrillation and pacing functions and play an important role in rescuing patients undergoing cardiac arrest and bradyarrhythmia (7). It has been shown that the defibrillation rate of temporary cardiac pacing can reach 100%, and the pacing success rate of patients with bradycardia is >80% (8). In addition, the energy required for biphasic defibrillation is minimal, the damage to the patient's myocardium is minor, and the time it takes for the heart to restore sinus rhythm after defibrillation is short. Temporary cardiac pacing therapy can stabilize the patients' hemodynamics and reduce the recurrence rate of cardiac arrest (9,10). In the present study, 12 patients were administered temporary pacemaker therapy, which controlled ventricular tachycardia and electrical storms in a timely manner, and maintained the vital signs of the patients. In addition to the temporary pacemaker, fluid replacement and antiarrhythmic treatment with amiodarone were continued. All 12 patients were cured.

The more common complications of temporary cardiac pacing include pacing disorder, arrhythmia, myocardial perforation, pericardial tamponade, pneumothorax, infection and puncture complications (such as subcutaneous hematoma and venous thrombosis) (11-13). In the present study, 1 patient experienced pacing dysfunction caused by temporary pacemaker displacement. Pacemaker malfunction was corrected after adjusting the position of the pacemaker.

Therefore, temporary cardiac pacing may be an effective treatment for malignant arrhythmias caused by severe acute aconitine poisoning. However, the present study has some limitations which are as follows: The present study involved a very small number of cases, and the data provided are thus limited. Therefore, further studies with larger sample sizes are required, and farther randomized, controlled multi-center studies are also warranted to confirm the current findings. There are reports on the implantation of a temporary external transvenous implantable cardioverter-defibrillator as a bridge for reimplantation following infected device extraction (14,15). Those temporarily implanted devices with a dual-coil ICD lead can deliver shocks (16); however, they were not used in the present study. A temporary pacemaker cannot treat ventricular tachycardia/ventricular fibrillation with biphasic cardioversion or defibrillation. Advanced life support/advanced cardiac life support for hemodynamically unstable patients with malignant ventricular arrhythmias is always electrical cardioversion, or CPR and defibrillation in case of cardiac arrest. Temporary pacemakers may be useful for maintaining stable cardiac output after effective defibrillation/cardioversion or they can provide overdrive pacing that can terminate fast ventricular rhythms. However, they are not a treatment of choice for patients with an electrical storm where external defibrillation/cardioversion is a priority. A more in-depth analysis of the side-effects which may occur prior to large-scale implementation is thus essential.

Acknowledgements

Not applicable.

Funding

Funding: The present study received funding via a grant provided by the National Natural Science Foundation (grant no. 81960350).

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Authors' contributions

FC, LH and MWL made contributions to the acquisition and analysis of the data. FC and MWL contributed to the interpretation of the data. FC, LH and MWL made contributions to the conception and design of the study and drafted the manuscript. LH and MWL prepared Fig. 1, Fig. 2 and Fig. 3. FC, LH and MWL confirm the authenticity of all the raw data. All authors have read and approved the final manuscript.

Ethics approval and consent to participate

The present study was approved by the First Affiliated Hospital of Kunming Medical University Research Ethics Boards. Written informed consent for participation in the study was obtained from the parent or relative of participants.

Patient consent for publication

Consent for patient data to be published was provided by the patients themselves or their relatives.

Competing interests

The authors declare that they have no competing interests.

References

1 

Tomlinson B, Chan TY, Chan JC and Critchley JA: Herb-induced aconitine poisoning. Lancet. 341:370–371. 1993.PubMed/NCBI View Article : Google Scholar

2 

Jeon SY, Jeong W, Park JS, You Y, Ahn HJ, Kim S, Kim D, Park D, Chang H and Kim SW: Clinical relationship between blood concentration and clinical symptoms in aconitine intoxication. Am J Emerg Med. 40:184–187. 2021.PubMed/NCBI View Article : Google Scholar

3 

Jin R, Wang YG and Zhang B: Establishment and verification of risk assessment scale for clinical safety medication of aconitine. Zhongguo Zhong Yao Za Zhi. 43:222–226. 2018.PubMed/NCBI View Article : Google Scholar : (In Chinese).

4 

Cho YS, Choi HW, Chun BJ, Moon JM and Na JY: Quantitative analysis of aconitine in body fluids in a case of aconitine poisoning. Forensic Sci Med Pathol. 16:330–334. 2020.PubMed/NCBI View Article : Google Scholar

5 

Chatzidou S, Kontogiannis C, Tsilimigras DI, Georgiopoulos G, Kosmopoulos M, Papadopoulou E, Vasilopoulos G and Rokas S: Propranolol versus metoprolol for treatment of electrical storm in patients with implantable cardioverter-defibrillator. J Am Coll Cardiol. 71:1897–1906. 2018.PubMed/NCBI View Article : Google Scholar

6 

Pan MC, Zhou XW, Liu Y, Wang YN, Qiu XG, Wu SF and Liu Q: Research progress on the molecular mechanisms of toxicology of ethanol-aconitine induced arrhythmia. Fa Yi Xue Za Zhi. 36:115–119. 2020.PubMed/NCBI View Article : Google Scholar : (In English, Chinese).

7 

He WW and Chu YJ: Clinical investigation of temporary heart pacemaker for emergency treatment of patients with acute and severe cardiovascular diseases. Zhonghua Yi Xue Za Zhi. 96:2644–2647. 2016.PubMed/NCBI View Article : Google Scholar : (In Chinese).

8 

Sert A, Aypar E, Odabas D and Aygul MU: Temporary cardiac pacemaker in the treatment of junctional rhythm and hypotension due to imipramine intoxication. Pediatr Cardiol. 32:521–524. 2011.PubMed/NCBI View Article : Google Scholar

9 

Lee MS, Nguyen H and Shlofmitz R: Incidence of bradycardia and outcomes of patients who underwent orbital atherecto-my without a temporary pacemaker. J Invasive Cardiol. 29:59–62. 2017.PubMed/NCBI

10 

Kowey PR, Mullan DF and Wetstein L: Pacemaker therapy. Surg Clin North Am. 65:595–611. 1985.PubMed/NCBI View Article : Google Scholar

11 

Metkus TS, Schulman SP, Marine JE and Eid SM: Complications and outcomes of temporary transvenous pacing: An analysis of > 360,000 patients from the national inpatient sample. Chest. 155:749–757. 2019.PubMed/NCBI View Article : Google Scholar

12 

Zhou X, Ze F, Li D, Li X and Wang B: Outcomes of temporary pacing using active fixation leads and externalized permanent pacemakers in patients with cardiovascular implantable electronic device infection and pacemaker dependency. J Cardiovasc Electrophysiol. 32:3051–3056. 2021.PubMed/NCBI View Article : Google Scholar

13 

Buttigieg J, Asciak R and Azzopardi CM: Pacemaker lead-associated thrombosis in cardiac resynchronisation therapy. BMJ Case Rep. 2015(bcr2015210314)2015.PubMed/NCBI View Article : Google Scholar

14 

Dębski M, Ząbek A, Boczar K, Urbańczyk-Zawadzka M, Lelakowski J and Małecka B: Temporary external implantable cardioverter-defibrillator as a bridge to reimplantation after infected device extraction. J Arrhythm. 34:77–80. 2017.PubMed/NCBI View Article : Google Scholar

15 

Da Rosa MR, Sapp JL, Howlett JG, Falkenham A and Légaré JF: Implantable cardioverter-defibrillator implantation as a bridge to cardiac transplantation. J Heart Lung Transpl. 26:1336–1339. 2007.PubMed/NCBI View Article : Google Scholar

16 

Kumar P, Baker M and Gehi AK: Comparison of single-coil and dual-coil implantable defibrillators: A meta-analysis. JACC Clin Electrophysiol. 3:12–19. 2017.PubMed/NCBI View Article : Google Scholar

Related Articles

Journal Cover

January-February 2022
Volume 4 Issue 1

Print ISSN: 2632-2900
Online ISSN:2632-2919

Sign up for eToc alerts

Recommend to Library

Copy and paste a formatted citation
x
Spandidos Publications style
Cheng F, Hao L and Liu M: Clinical effect of a temporary pacemaker on an electrical storm induced by severe acute aconitine poisoning. World Acad Sci J 4: 7, 2022
APA
Cheng, F., Hao, L., & Liu, M. (2022). Clinical effect of a temporary pacemaker on an electrical storm induced by severe acute aconitine poisoning. World Academy of Sciences Journal, 4, 7. https://doi.org/10.3892/wasj.2022.142
MLA
Cheng, F., Hao, L., Liu, M."Clinical effect of a temporary pacemaker on an electrical storm induced by severe acute aconitine poisoning". World Academy of Sciences Journal 4.1 (2022): 7.
Chicago
Cheng, F., Hao, L., Liu, M."Clinical effect of a temporary pacemaker on an electrical storm induced by severe acute aconitine poisoning". World Academy of Sciences Journal 4, no. 1 (2022): 7. https://doi.org/10.3892/wasj.2022.142